TEST INST (S)
CALIBRATION CHECKLIST
PROC. NO.
Beckman Instruments 2000 Ion Analyzer
MFG. MODEL
SER. NO.
PROCEDURE
STEP
NO.
(1)
FUNCTION TESTED
(2)
NOMINAL
(3)
MEASURED VALUES
FIRST RUN
(4)
SECOND RUN
(5)
OUT
OF
TOL
(6)
CALIBRATION TOLERANCES
(7)
NA 17–20SC–42
Page 1 of 3
3.1 Test Control Settings
TI control(s):
3.1.4 Depress pH, EXP, and STBY
pushbuttons
TI TEMPERATURE and
SLOPE controls to OFF
MIDSCALE VALUE set to 7
3.1.5 Release STBY pushbutton
adjust needle to 0 on the
+
0.7 scale 0 ck( ) NA
4.1 Millivolt Indication Tests
4.1.3 Depress +mV and ABS
pushbuttons
Release EXP pushbuttons
Zero Indication (mV) (mV)
0–2to+2
4.1.4 mV Range (mV) (mV)
1300 1298 to 1302
1100 1098 to 1102
900 898 to 902
700 698 to 702
500 498 to 502
300 298 to 302
100 98 to 102
Depress TI – mV pushbutton
–100 –98 to –102
–300 –298 to –302
–500 –498 to –502
–700 –698 to –702
–900 –898 to –902
NAVAIR 17–20SC–42
15